WGM - 28 November 2007 - Edition 01
PAT700 Total Organic Carbon Analyzer – Specifications Page 141
Anatel Operator Manual
Appendix A: Specifications
PAT700 specifications are subject to change without notice.
TOC specifications
Operating range 0.5 to 2,000 ppb as carbon
Display resolution 0.1 ppb
Accuracy ±1 ppb or 5%, whichever is greater
Repeatability ±0.3 ppb or ±1%, whichever is greater
Limit of detection 0.5 ppb
Conductivity and resistivity specifications
Conductivity range (TOC) 0.05 to 150 µS/cm (@ 25 °C)
Conductivity display resolution 0.01 µS/cm
Conductivity accuracy ±1% over full range (uncompensated)
Resistivity range 0.2 to 18 M
Ω-cm (@25 °C)
Resistivity display resolution 0.01 over full range
Available resistivity modes Temperature compensated to 25 °C, or uncompensated
Temperature specifications
Ambient operating range 10 to 40° C (50 to 104 °F)
Display resolution 0.1 over full range
Measurement accuracy ±0.4°C
Sample water range 5 to 95 °C (41 to 203 °F)