Page 12 Installation – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
WGM - 28 November 2007 - Edition 01
Mounting to a wall
Follow these guidelines and refer to Figure 2-1, Figure 2-2,and Figure 2-3 to mount the
transmitter to a wall or other flat, rigid surface:
• Use 5/16-inch (8 mm) diameter screws or bolts and nuts to mount the analyzer to a
wall or other flat, rigid surface. Use hardware that can withstand the process
environment. Hach Ultra does not supply these screws, bolts, or nuts.
• To minimize stress on the analyzer mounting bracket, secure all mounting bolts to the
same structure, which should be flat and should not vibrate or move excessively. Do
not secure bolts to separate girders, beams, or wall studs that can move
independently.
Figure 2-3 PAT700 wall mount
Mounting to dual instrument poles
Follow these guidelines and refer to Figure 2-1 and Figure 2-4 to mount the analyzer to
instrument poles:
• Mount the analyzer mounting bracket to two instrument poles that are attached to a
common surface. Do not secure to two instrument poles that could move
independently.
• The instrument poles should extend at least 10 inches (254 mm) from a common rigid
base and should be no larger than 2 inches (50.8 mm) in diameter.