Page 20 Installation – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
WGM - 28 November 2007 - Edition 01
Discrete inputs
The analyzer has two discrete inputs. Wiring connects to terminals on connector block J24, as
listed in Table 2-4.
Discrete input 1 initiates TOC analysis. When the output state switches from high to
low, the analyzer stops the current operation and runs a single TOC analysis. After the
analysis is complete, the analyzer returns to the mode defined by digital input 2. During
the TOC analysis, the analyzer ignores all subsequent trigger values.
Discrete input 2 switches the analyzer modes. If digital output 2 is in high state, the
analyzer is in TOC mode. If digital output 2 is in low state, the analyzer runs in flow with
conductivity mode.
Discrete outputs
The analyzer has four discrete outputs. Wiring connects to terminals on connector block J22,
as listed in Table 2-5.
Discrete output 1 is a TOC alarm. The output reports the TOC level as above (low
state) or below (high state) the user-specified alarm limit.
Discrete output 2 is an uncompensated conductivity alarm. A low state indicates the
conductivity level is above the alarm limit. A high state indicates the conductivity level
is below the alarm limit.
Table 2-3
J17 connector block: 4-20 mA output wiring terminal designations
J17 Block Terminal Description Variable Designation
1 4-20 mA source output for analog output 1 (+) TOC AO1+
2 4-20 mA sink output for analog output 1 (–) TOC AO1-
3 4-20 mA source output for analog output 2 (+) Conductivity or resistivity AO2+
4 4-20 mA sink output for analog output 2 (–) Conductivity or resistivity AO2-
5 4-20 mA source output for analog output 3 (+) Temperature AO3+
6 4-20 mA sink output for analog output 3 (–) Temperature AO3-
Table 2-4 J24 connector block: Discrete input wiring terminal designations
J24 Block Terminal Description Variable Designation
1 Common connection for external sourcing (+) Common DI1/2+
2 Digital input 1 voltage source return input (–) Initiates TOC analysis DI1–
3 Digital input 2 voltage source return input (–) Analyzer mode DI2–
4 12 VDC power supply +
Connect only if output device requires
power supply
Output power supply +12V
5 12 VDC power supply –
Connect only if output device requires
power supply
Output power supply GND