EasyManua.ls Logo

Hach ANATEL PAT700 - I;O Wiring

Hach ANATEL PAT700
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WGM - 28 November 2007 - Edition 01
PAT700 Total Organic Carbon Analyzer – Installation Page 23
Anatel Operator Manual
I/O wiring
The analyzer has three 4-20 mA outputs, two discrete inputs, and four discrete outputs.
4-20 mA outputs
The analyzer has three 4-20 mA outputs, as listed in Table 2-7.
Analog output 1 represents TOC (total organic carbon).
Analog output 2 represents conductivity or resistivity (user selected).
Analog output 3 represents temperature in °F or °C (user selected).
Discrete inputs
The analyzer has two discrete inputs, as listed in Table 2-8.
Discrete input 1 initiates TOC analysis. When the output state switches from high to
low, the analyzer stops the current operation and runs a single TOC analysis. After the
analysis is complete, the analyzer returns to the mode defined by digital input 2. During
the TOC analysis, the analyzer ignores all subsequent trigger values.
Discrete input 2 switches the analyzer modes. If digital output 2 is in high state, the
analyzer is in TOC mode. If digital output 2 is in low state, the analyzer runs in flow with
conductivity mode.
Table 2-7 4-20 mA output wiring terminal designations
Wiring connector Terminal Description Variable Designation
1 4-20 mA source output for analog output 1 (+) TOC AO1+
2 4-20 mA sink output for analog output 1 (–) TOC AO1–
3 4-20 mA source output for analog output 2 (+) Conductivity or
resistivity
AO2+
4 4-20 mA sink output for analog output 2 (–) Conductivity or
resistivity
AO2–
5 4-20 mA source output for analog output 3 (+) Temperature AO3+
6 4-20 mA sink output for analog output 3 (–) Temperature AO3–
Table 2-8 Discrete input wiring terminal designations
Wiring connector Terminal Description Variable Designation
1 Common connection for external sourcing (+) Common DI1/2+
2 Digital input 1 voltage source return input (–) Initiates TOC analysis DI1–
3 Digital input 2 voltage source return input (–) Analyzer mode DI2–
4 12 VDC power supply +
Connect only if output device requires power
supply
Output power supply +12V
5 12 VDC power supply –
Connect only if output device requires power
supply
Output power supply GND