WGM - 28 November 2007 - Edition 01
PAT700 Total Organic Carbon Analyzer – Installation Page 33
Anatel Operator Manual
Example:
Read command: PV
Response:
Diagnostic commands
Example:
Read command: “RM”
Read format: “Mode State”
Response: “0 0”
“PV”
Show TOC validation history
“RM”
Get analyzer mode and state
Mode State
0 = Offline 0 = Idle
1 = Online TOC 1 = Start sample
2 = Online conductivity 2= Calibrate
3 = Digital TOC 3 = Start flushing
4 = Manual TOC 4 = Flushing
5 = Manual bottle 5 = Wait 15 seconds
6 = Manual clean 6 = Reading TOC
7 = Manual diagnostic 8 = Analysis done
8 = Standby 9 = Initialization error
10 = Error