EasyManua.ls Logo

Hach ANATEL PAT700 - Log View Tab

Hach ANATEL PAT700
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WGM - 28 November 2007 - Edition 01
PAT700 Total Organic Carbon Analyzer – Startup Page 39
Anatel Operator Manual
Temperature is the last temperature reading in “Temperature: X.X °C” format or
“Temperature: X.X °F” format, depending on the temperature units selected in the
setup dialog box.
Sample time is the time of the start of the last reading in 12-hour “Sample Time:
hh:mm:ss” format.
Trend is the TOC trend over the last hour in “Trend: +/-X.X ppb/hr” format.
Curve type is the curve type for the last TOC reading in “Curve Type: X” format, where
X can be one of four types (1, 2, 3, or 5).
Date and time (top of screen) displays the current date and time in ISO format. Depending on
the current time format setting, the date and time is in 24-hour format (yyyy-mm-dd hh:mm:ss)
or 12-hour format (yyyy-mm-dd hh:mm:ss AM/PM). The date and time are configured in the
setup dialog box.
Bottles icon (lower left corner of screen) displays the current state of the bottles in the bottle
bay. There are three states for this indicator (no bottles loaded, bottles loaded, or bottles
empty). Touching the icon takes you to the bottle mode dialog box.
Status indicator (lower right corner of screen) displays the status of the two UV lamps.
Green indicates the UV lamp is good.
Yellow indicates the UV lamp is marginal and should be replaced.
Red indicates the UV lamp has failed and must be replaced.
Process animation (right side of screen) displays the current state of the analysis as an
animated graphic. There are five possible states: water flowing through the cell, water not
flowing through the cell, analyzing, new alarm condition, and acknowledged alarm condition.
Analysis results are displayed until a new analysis is complete. It displays the last readings for
conductivity, temperature, and the time of the last reading.
Log view tab
The log view tab displays all analysis readings or events as they would appear on the printout.
The display include all data, up to the last 72 hours, since the last power-up. See Figure 3-4.
Figure 3-4 Home screen, log view tab

Table of Contents

Other manuals for Hach ANATEL PAT700