Page 53 Setup – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
WGM - 28 November 2007 - Edition 01
Static subnet mask text box sets the subnet mask for the analyzer. You may specify a value
of 0 to 255 for each octet. The default value is 255.255.255.0.
Static gateway text box sets the default gateway for the analyzer. You may specify a value
of 0 to 255 for each octet. The default value is blank.
DNS server text box sets the DNS server address for the analyzer. You may specify a value
of 0 to 255 for each octet. The default value is blank.
WINS server text box sets the WINS server address for the analyzer. You may specify a
value of 0 to 255 for each octet. The default value is blank.
DHCP check box allows you to enable or disable DHCP. When enabled, the TCP/IP address,
subnet mask, and default gateway entries are disabled. The default setting is enabled.
Return icon returns you to the setup dialog box and saves the changes. Any changes are
noted in the audit trail. All changes take effect immediately.
5.5 Alarms
The alarms icon allows you to enable or disable TOC and conductivity alarms for the system.
Figure 5-8 Alarm dialog box
TOC upper limit text box allows you to set the alarm upper limit for online TOC
measurement. You may enter a value from 0 to 2000. The default value is 500. The value
applies only to online TOC analysis, not to bottle tests or grab sample analysis.
TOC enabled checkbox allows you to enable or disable the online TOC alarm. If enabled,
any online TOC reading that exceeds the alarm upper limit triggers a TOC alarm. If disabled,
no TOC alarms are triggered. The default setting is enabled.
Conductivity enabled checkbox allows you to enable or disable the conductivity alarm. If
enabled, any conductivity reading that exceeds the value in Table 5-1 for the current
temperature triggers a conductivity alarm. If disabled, no conductivity alarms are triggered.
The default setting is enabled.
Alarms are reported for conductivity excursions only in the uncompensated mode and require
that uncompensated conductivity be selected.