Page 58 Setup – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
WGM - 28 November 2007 - Edition 01
TOC tab
Figure 5-10 Printer dialog box, TOC tab
Disabled, continuous, and percent change radio buttons determine how often an online
TOC reading will be printed. You may select disabled, continuous, or percent change.
Selecting disabled disables printing of online TOC readings.
Selecting continuous (default) causes a TOC printout to occur at the end of each
analysis cycle and disables the percent change text box.
Selecting percent change causes a TOC printout to occur only when the change in
online TOC readings between successive analyses exceeds the value specified in the
percent change box and enables the percent change text box.
Limit text box allows you to enter the percentage change in online TOC readings that must
occur between successive analyses before a printout occurs. You may select a value from 1
to 100%. The default value is 1.
Return icon returns you to the setup dialog box and saves the changes. Any changes are
noted in the audit trail.