Page 67 Setup – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
WGM - 28 November 2007 - Edition 01
User must manually accept results and auto accept results radio buttons determine
whether the outcome of the test are manually or automatically accepted. You may select user
must manually accept results (default) or auto accept results and return online.
If user must manually accept results is selected, the results of the calibration will not be
used until you accept them.
If auto accept results and return online is selected, the slope and R
2
limit options are
enabled. If the calibration is accepted, the new calibration factors are used.
Return to online mode checkbox tells the analyzer whether or not to return to online mode
following the TOC calibration. You may enable or disable (default) this option. When enabled,
the analyzer returns to online mode following the TOC calibration. When disabled, the
analyzer stays at the bottle mode dialog box. All valves are closed, the pump is off, and the
lamp is off.
Slope limit text box allows you to specify the acceptable limit of the slope for the calibration
to pass. You may enter a value from 1 to 15%. The default value is 10%.
R2 limit text box allows you to specify the acceptable limit for the determination coefficient.
You may enter a value from 0.9000 to 0.9999. The default value is 0.9900.
Return icon returns you to the setup dialog box and saves any changes. Any changes are
noted in the audit trail.
Conductivity calibration tab
Figure 5-21 Bottle test setup dialog box, conductivity calibration tab
User must manually accept results and auto accept results radio buttons determine
whether the outcome of the test are manually or automatically accepted. You may select user
must manually accept results (default) or auto accept results and return online.
If user must manually accept results is selected, the results of the calibration will not be
used until you accept them.
If auto accept results and return online is selected, the slope limit options are enabled.
If the calibration is auto accepted, the new calibration factors are used.