• Summary text lists the results of the test. The results appear as they would appear
on the printed report.
• Export icon takes you to the export dialog box, where you can save the results of
the test to a USB flash drive.
• Checkmark icon accepts the calibration, if the values passed, and takes you to the
unload/replace bottles dialog box.
• Cancel icon rejects the calibration results and takes you to a dialog box that asks if
you want to rerun the test. If you touch the forward icon, the analyzer returns you to
the beginning of the test. All previous data are discarded.
• Auto accept. If you have selected to auto accept the results and return online, the
results are accepted and sent to the printer, and the analyzer returns to the home
screen. If you have selected manually accept the results and return online, the
analyzer returns to the home screen immediately following the completion of the
test. If you have selected manually accept the results and remain offline, the test
summary dialog box appears.
12) When the unload/replace bottles dialog box appears, wait for the bottles to
backflush and equilibrate pressure, follow the instructions for unloading or replacing
bottles, then press the done icon. See Figure 6-11.
Figure 6-11 Unload/replace bottles dialog box
• Unlock icon energizes the bottle bay door lock for five seconds, allowing you to
open the door.
• Done icon returns you to the bottle mode dialog box.
13) If the test failed or you have rejected the calibration, touch the cancel icon, then
touch the checkmark icon to run the calibration again (see Figure 6-12) or the
cancel icon to cancel the run.