Page 86 Bottle Mode – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
WGM - 28 November 2007 - Edition 01
7) Touch the << icon to open the sliding toolbar, then touch the bottles icon in the
toolbar, or touch the bottles icon in the lower left corner of the screen.
8) Touch the run standards button in the bottle mode dialog box. See Figure 6-1,
page 76.
9) Touch the conductivity calibration button in the run standards dialog box. See
Figure 6-2, page 77.
10) The OASIS door will unlock for five seconds. If the door is not opened within five
seconds, the unlock icon can be used to unlock the door manually.
11) Install the standards bottle following the instructions as shown in the load bottles
dialog box and ensuring that you place the standards bottle in the correct position.
See Figure 6-13. From left:
Bottle 1 = empty
Bottle 2 = 100 µS potassium chloride conductivity test standard
Bottle 3 = empty
Bottle 4 = empty
Figure 6-13 Load bottles dialog box, conductivity calibration
Bottle positions graphics show you the order in which to install the bottles.
Unlock icon energizes the bottle bay door lock for five seconds, allowing you to
open the door.
Back icon takes you to the run standards dialog box.
Forward icon takes you to the review test options dialog box.
Cancel icon returns you to the previous dialog box. All changes made in any of the
dialog boxes are lost.
12) Close the bottle bay door.
13) Touch the forward icon. If bottles have been properly loaded and the RFID tags can
be read, the review test setup dialog box appears. See Figure 6-6, page 81.
Otherwise, the enter bottle information dialog box appears. See Figure 6-7,
page 81.