WGM - 28 November 2007 - Edition 01
PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 87
Anatel Operator Manual
14) If the selection has been made to use the conductivity meter verification resistor,
install the resistor onto the analyzers 9-pin printer/calibration port, as illustrated in
Figure 6-14.
If the calibration is being performed on a schedule, the conductivity meter test
dialog box does not appear, and the calibration occurs automatically.
Otherwise, the conductivity meter test dialog box appears when the analyzer
passes the conductivity meter test. See Figure 6-15.
Figure 6-14 Verification resistor installed on printer/calibration port
Figure 6-15 Conductivity meter test dialog box
Verification resistor