Page 105
Bottle Mode
Bottle positions graphics show you the order in which to install the bottles.
Unlock icon energizes the bottle bay door lock for five seconds, allowing you to
open the door.
Back icon takes you to the run standards dialog box.
Forward icon takes you to the review test options dialog box.
Cancel icon returns you to the run standards dialog box. All changes made in any
of the dialog boxes are lost.
7. Close the bottle bay door.
8. Touch the forward icon. If bottles have been properly loaded and the RFID tags can
be read, the review test setup dialog box appears. See Figure 72, page 93.
Otherwise, the enter bottle information dialog box appears. See Figure 73, page 93.
9. When the test running dialog box appears, the analyzer begins running the test.
Three samples are drawn from each bottle. After all the repetitions are complete,
the analyzer repeats the process for the rest of the bottles. All reported values
must be normalized to 500 ppb using the following formula.
Once all the bottles have been analyzed, the analyzer compares the average
reading of the sucrose standard to the average reading of the 1,4 benzoquinone.
If the two values are within ±15% of each other, the test passes. The analyzer
then displays the test summary dialog box. The following formulas are used to
calculate the response efficiency.
Figure 85 Load bottles dialog box, system suitability test
r
n
r
raw
500
C of A value
--------------------------------
⎝⎠
⎛⎞
=
r
e
100
SR
LR
--------
⎝⎠
⎛⎞
×=