Page 123
Alarms
149 UnableToStartPlumbingTest Analyzer cannot start plumbing test.
150 UnableToPrintDataLog Analyzer cannot print data log.
151 LampExtinct
• Analyzer detects that current lamp has failed.
• Replace UV lamp.
153 Bottle1MarkedAsUsed
• Bottle installed in location 1 is marked as used.
• Replace bottle with a valid standard.
154 Bottle2MarkedAsUsed
• Bottle installed in location 2 is marked as used.
• Replace bottle with a valid standard.
155 Bottle3MarkedAsUsed
• Bottle installed in location 3 is marked as used.
• Replace bottle with a valid standard.
156 Bottle4MarkedAsUsed
• Bottle installed in location 4 is marked as used.
• Replace bottle with a valid standard.
157 ExcursionBottle1NotEmpty
• Excursion bottle 1 is not empty.
• Remove excursion bottle 1 and replace with empty bottle.
158 ExcursionBottle2NotEmpty
• Excursion bottle 2 is not empty.
• Remove excursion bottle 2 and replace with empty bottle.
159 ExcursionBottle3NotEmpty
• Excursion bottle 3 is not empty.
• Remove excursion bottle 3 and replace with empty bottle.
160 ExcursionBottle4NotEmpty
• Excursion bottle 4 is not empty.
• Remove excursion bottle 4 and replace with empty bottle.
172 RougeDetected
• One or more of your 500 ppb standards took longer than
expected to oxidize.
• Clean the cell.
165 LampSwitch
• Analyzer has switched UV lamps because current lamp was
marginal or failed.
• Replace UV lamp.
178 ExcursionCaptured
• An excursion sample was captured in excursion bottle.
• Remove sample from analyzer.
180 ConductivityMeterAlarm Analyzer has not passed conductivity meter test.
181 BottleTestCancelled Bottle test was cancelled by the user.
Table 16 Warning alarm messages
Code Message Alarm condition