TOC specifications
Operating range 0.5 to 2,000 ppb as carbon
Conductivity Range for TOC
0.2 µS/cm for all waters, 1.0 µS/cm for all neutral waters, 5 µS/cm for water with
CO
2
as the sole conductive species.
Display resolution 0.1 ppb
Accuracy ±1 ppb or 5%, whichever is greater
Repeatability ±0.3 ppb or ±1%, whichever is greater
Limit of detection 0.5 ppb
Conductivity and resistivity specifications
Conductivity range 0.05 to 150 µS/cm (@ 25 °C)
Conductivity display resolution 0.01 µS/cm
Conductivity accuracy ±1% over full range (uncompensated)
Resistivity range 0.2 to 18 M
Ω-cm (@25 °C)
Resistivity display resolution 0.01 over full range
Available resistivity modes Temperature compensated to 25 °C, or uncompensated
Temperature specifications
Ambient operating range 10 to 40° C (50 to 104 °F)
Display resolution 0.1 over full range
Measurement accuracy ±0.4°C
Sample water range 5 to 95 °C (41 to 203 °F)
Physical specifications
UV lamps Two, with UV Detect technology
Interface/display Color touch screen
Maximum altitude 4,000 m (13,125 ft)
User I/O wiring Three 3/4-inch conduit openings or quick disconnect fittings
Standards system Onboard, automated standards introduction system (OASIS)
Dimensions
59.7 cm wide x 22.9 cm deep x 25.4 cm high
(23.5” wide x 9” deep x 10” high)
Weight 13.6 kg (30 lbs)
Sample inlet flow rate range 60 mL/min to 300 mL/min
Sample inlet pressure range 10 to 100 psi
Electronics enclosure rating
IP56 conduit version
IP46 quick connect version
Humidity rating 5 to 95%
CE compliance EN61010-1 and EN 61326
Safety rating ETL, conforming to UL 61010-1 and CSA 22.2 No. 61010-1
Protection and overvoltage Category II
Pollution degree 2, IEC 61010-1
Power input 100 to 230 ±10% VAC, 65 W, 50/60 Hz