EasyManua.ls Logo

Hach ANATEL PAT700 - Parameter Setup Command

Hach ANATEL PAT700
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 38
Installation
2.8.6 Parameter setup command
Example:
Read command: “SY”
Response:
“7 25 2007 20 33 31”
Write format:
“SY mm DD YYYY HH MM SS”
“SY 7 30 2007 12 30 25”
1. Flush time (HH:mm:SS)
2. Cycle time (HH:mm:SS)
3. TOC alarm limit (ppb)
4. Pump sampling (0 = Disabled, 1 = Enabled)
5. Sensor name (1 to 13 characters)
6. Idle mode (0 = No flow, 1 = Flow with conductivity, 2 = Flow without conductivity)
7. Always 0
8. Resistivity/conductivity units (0 = Mohm-cm, 1=
μS/cm C, 2 = Mohm-cm U, 3 = μS U)
9. Always 0
10. TOC mode print strategy (0 = Continuous, 1 = Percent change, 2 = Disabled)
11. Conductivity print strategy (0 = Timed, 1 = Percent change, 2 = Disabled)
12. Active alarm type (0 = TOC only, 2 = TOC and uncompensated conductivity)
13. Analog output type (0 for 4-20 mA, 1 for 0-20 mA)
14. Analog output on error (0 = Minimum [2mA], 1 = Last output, 2 = Maximum [22mA])
15. TOC percent change between printouts (1-99)
16. Conductivity percent change between printouts (1-99)
17. Analog zero-scale TOC (ppb)
18. Analog full-scale TOC (ppb)
19. User TOC calibration slope
20. Digital control mode (0 = Disabled, 1 = Enabled)
21. Always 0 (data and audit Log)
22. User conductivity calibration slope
“SY” Set/get system date and time
“CA”
Disable automatic result printouts
for serial port
Disables the automatic printout of results for the serial port.
“HR” Set/get global settings
Sets or gets global settings. Compatible with or without user
conductivity calibration slope parameter.

Table of Contents

Other manuals for Hach ANATEL PAT700

Related product manuals