Page 88
Bottle Mode
6.1.5 System suitability protocol
Perform a system suitability test according to the guidelines established in USP Method
<643> using prepackaged 500 ppb sucrose and 500 ppb 1,4-benzoquinone standards.
6.2 Accessing bottle mode
To access bottle mode, follow either of these procedures:
Touch the << icon to open the sliding toolbar, then touch the bottles icon in the toolbar,
or
Touch the bottles icon in the lower left corner of the home screen.
Note: If security is enabled, you must be signed on the analyzer to access the bottle mode. Bottle
tests can not be cancelled or accepted/rejected when security is enabled and a user is logged off the
analyzer.
6.3 Bottle mode dialog box
The bottle mode dialog box allows you to perform procedures using the bottles in the
bottle bay. You may run calibrations, validations, or system suitability tests, analyze
samples from a bottle (grab sample), or install an excursion bottle to capture a sample
when a TOC measurement exceeds an alarm limit, a conductivity alarm occurs, specific
alarm conditions occurs, or manually. See Figure 67.
Run standards button takes you to the run standards dialog box. See Run standards,
page 89.
Grab sample button takes you to the grab sample load bottles dialog box. See Grab
sample, page 106.
Excursion mode button takes you to the excursion mode dialog box. See Excursion
mode, page 110.
Unload bottles button takes you to the unload/replace bottles dialog box. See
Figure 78, page 97.
Figure 67 Bottle mode dialog box