EasyManua.ls Logo

Hach AS950 - Section 5 Installation; 5.1 Confined space precautions; 5.2 Site installation guidelines

Hach AS950
172 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 5 Installation
D A N G E R
Multiple hazards. Only qualified personnel must conduct the tasks described in this
section of the document.
5.1 Confined space precautions
D A N G E R
Explosion hazard. Training in pre-entry testing, ventilation, entry procedures,
evacuation/rescue procedures and safety work practices is necessary before entering
confined spaces.
The information that follows is supplied to help users understand the dangers and risks that are
associated with entry into confined spaces.
On April 15, 1993, OSHA's final ruling on CFR 1910.146, Permit Required Confined Spaces, became
law. This standard directly affects more than 250,000 industrial sites in the United States and was
created to protect the health and safety of workers in confined spaces.
Definition of a confined space:
A confined space is any location or enclosure that has (or has the immediate potential for) one or
more of the following conditions:
An atmosphere with an oxygen concentration that is less than 19.5% or more than 23.5% and/or a
hydrogen sulfide (H
2
S) concentration that is more than 10 ppm.
An atmosphere that can be flammable or explosive due to gases, vapors, mists, dusts or fibers.
Toxic materials which upon contact or inhalation can cause injury, impairment of health or death.
Confined spaces are not designed for human occupancy. Confined spaces have a restricted entry
and contain known or potential hazards. Examples of confined spaces include manholes, stacks,
pipes, vats, switch vaults and other similar locations.
Standard safety procedures must always be obeyed before entry into confined spaces and/or
locations where hazardous gases, vapors, mists, dusts or fibers can be present. Before entry into a
confined space, find and read all procedures that are related to confined space entry.
5.2 Site installation guidelines
D A N G E R
Explosion hazard. The instrument is not approved for installation in hazardous locations.
This instrument is rated for an altitude of 2000 m (6562 ft) maximum. Although the use of this
equipment above the 2000 m altitude does not show any substantial safety concern, the
manufacturer recommends that users with concerns contact technical support.
Refer to the guidelines that follow for the site location evaluation.
Obey all the safety precautions if the sampler is installed in a confined space. Refer to Confined
space precautions on page 10.
Make sure that the temperature at the location is in the specification range. Refer to Specifications
on page 3.
Install the sampler on a level surface or hang the sampler with the suspension harness, the
support bracket or the spanner bar. Refer to Install the sampler in a manhole on page 11 and to
the applicable installation documentation.
Install the sampler as near the sample source as possible. Refer to Plumb the sampler
on page 13.