Total Oxidants
Total Oxidants
Page 122
50081_TotalOxidants.fm
17. Use the navigation keys
to select two line segments,
one on each side of the
titration end point, by
positioning vertical lines
(cursors) at the beginning
and end of each segment.
Press the
CHECK MARK to
confirm changes or press
DEL to return to the Zoom
window.
Note: For more information on
manual EP determination, see
page 135.
18. Press the CHECK MARK
to accept the analysis and
proceed to Titration 2, or
press
DEL to return to the
previous window to adjust
the linear segments and
reanalyze the data.
Note: The analysis data may
be analyzed as many times as
needed from the Zoom Range
Determination window, in
manual and/or automatic
modes. The results are not
stored until accepted.
The AutoCAT stores all titration
results and then automatically
calculates each analyte value
after the procedure is complete.