Chlorite
Chlorite
Page 130
50081_Chlorite.fm
17. Select type of end point
determination.
Press
DEL for Manual.
Note: Automatic determination
attempts to identify linear
segments without user input is
not recommended for this
method.
18. Use the navigation keys
to select two line segments,
one on each side of the
titration end point, by
positioning vertical lines
(cursors) at the beginning
and end of each segment.
Press the
CHECK MARK to
confirm changes or press
DEL to return to the Zoom
window.
Note: For more information on
manual EP determination, see
page 135.
19. Press the CHECK MARK
to accept the analysis and
proceed to Titration 2, or
press
DEL to return to the
previous window to adjust
the linear segments and
reanalyze the data.