Iodine
Iodine
Page 58
50081_IodineCalibration.fm
1. Turn the instrument on
if necessary and access the
Titrants menu.
Press
4 to select
Titrant Library.
Note: Supervisor-mode is
required to access calibrations.
See Accessing Supervisor
Mode on page 33.
2. Highlight ID.
Press the
CHECK MARK
to change.
3. Highlight Iodine
28.2mN.
Press the
CHECK MARK
to validate.
4. Press 1 to select
Edit Titrant.
5. Press the CHECK MARK
to change the titrant
concentration input mode.
6. Highlight Calibrate.
Press the
CHECK MARK to
validate.
7. The titrant
concentration input mode is
now set to calibrate and the
calibration parameters are
now available by scrolling
this screen.
Press
ESC twice to return to
the Titrants menu.
8. Press 1 to select
Calibrate/Enter titrant conc.
Note: Titrant calibration is now
required before analyses may
be run as indicated by the
“Stormy” icon.
AutoCAT Calibration Method Iodine Titrant Calibration