Chlorine Dioxide Generator Yield
Chlorine Dioxide Generator Yield
50081_ChlorineDioxideGeneratorYield.fm Page 73
5. The Start Timer will
begin, allowing the
electrode to stabilize before
data is collected.
Once the electrode has
stabilized, the AutoCAT will
begin data acquisition and
plot the titration curve.
6. The Zoom Range
Determination window will
display the titration curve.
Select and move the cursors
to define the range of data
points for analysis.
7. Select the type of end
point determination.
Press
DEL for Manual.
Note: Automatic determination
attempts to identify linear
segments without user input
and is not recommended for
this method.
8. Using the navigation
keys, select two line
segments, one on each side
of the titration end point, by
positioning vertical lines
(cursors) at the beginning
and end of each segment.
Press the
CHECK MARK to
confirm changes.
Press
DEL to return to the
Zoom window.
Note: For more information on
manual EP determination, see
page 135.
9. Press the CHECK MARK
to accept the analysis and
proceed to Titration 3 or
press
DEL to return to the
previous window to adjust
the linear segments and