Chlorine Dioxide
Chlorine Dioxide
Page 88
50081_Chlorine_ChlorineDioxide.fm
9. The analysis parameters
that have been stored for the
current Sub-ID may be
edited at this time.
Highlight the appropriate
line and press the
CHECK MARK to change one
of the analysis parameters or
press 1 to confirm.
Note: Any changes made will
become the new default
parameters for this method
Sub-ID.
Note: These Sub-ID
parameters can be edited here
as 'run-time' settings, or from
the Main Menu as described in
Section 3.5.
Note: This step will be skipped
if autoscaling software is
enabled.
10. The Start Timer will
begin to allow the electrode
to stabilize before data is
collected.
After the electrode has
stabilized, the AutoCAT will
begin data acquisition and
plot the titration curve in real
time.
11. The Zoom range
determination window will
display the titration curve.
Select and move the cursors
to define the range of data
points for analysis.
Proceed to step 12.
12. Select type of end point
determination.
Press
DEL for Manual.
Note: Automatic determination
attempts to identify linear
segments without user input
but is not recommended for
this method.