Chlorine Dioxide
Chlorine Dioxide
50081_Chlorine_ChlorineDioxide.fm Page 89
13. Use the navigation
buttons to select two line
segments, one on each side
of the titration end point, by
positioning the cursors at
the beginning and end of
each segment (indicated by
arrows).
Press the
CHECK MARK to
confirm changes or press
DEL to return to the previous
window.
Note: For more information on
manual EP determination, see
page 135.
14. Press the CHECK MARK
to accept the analysis and
proceed to Titr. 4.
Press
DEL to return to the
previous window to adjust
the linear segments and
reanalyze the data.
Note: For more information on
manual EP determination, see
page 135.