Hach DR/850 - CYANURIC ACID Turbidimetric Method

Hach DR/850
460 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
165
Method 8139
CYANURIC ACID (7 to 55 mg/L) For water, pools and spas
Turbidimetric Method
1. Enter the stored
program number for
cyanuric acid.
Press:
PRGM
The display will show:
PRGM ?
Note: For most accurate
results, perform a Reagent
Blank Correction using
deionized water (see
Section 1).
2. Press: 24 ENTER
The display will show
mg/L, CYACD and the
ZERO icon.
3. Fill a sample cell
with 25 mL of sample
(the blank).
Note: Filtering is required
for highly turbid samples.
4. Place the blank into
the cell holder. Tightly
cover the sample cell
with the instrument cap.
5. Press: ZERO
The cursor will move to
the right, then the
display will show:
0 mg/L CYACD
Note: If Reagent Blank
Correction is on, the
display may flash “limit”.
See Section 1.
6. Fill another cell with
25 mL of sample.
7. Add the contents of
one Cyanuric Acid 2
Reagent Powder Pillow
(the prepared sample).
Swirl to mix.
8. Press TIMER
ENTER
A three-minute reaction
period will begin.
Note: A white turbidity will
form if cyanuric acid is
present.
Note: Accuracy is not
affected by undissolved
powder.