EasyManua.ls Logo

Hach DR2400 - Page 209

Hach DR2400
798 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DR/2400
Chlorine, Total
ChlorineTot_PT2_ULow_DPD_Eng_Ody.fm Page 1 of 10
• Analyze samples immediately. Chlorine is not stable in an aqueous solution.
• A reagent blank value for a combined lot of indicator/buffer reagent solutions should be determined at least once a day. See
Determining the Reagent Blank Value on page 4. To store the reagent blank value, see Creating a User Program Based on a
Hach Program in the instrument manual.
• The reagent blank value is normally less than 5 µg/L. If the value is greater than 5 µg/L, an interfering substance may be present
in the blanking water or the DPD Indicator may be degrading. If there is doubt about the reagents, repeat the reagent blank
determination using chlorine-demand-free water for the sample. Blanks up to 5 µg/L may be used.
• Use a new filter for each test. Using an unspecified filter may give low analysis results or inability to filter the required volume.
• The ampules contain more than 1.0 mL of solution for ease of transfer. Discard excess reagent in the ampule.
• Install the Pour-Thru Cell into the adapter so that the long pathlength (1-in./25-mm) is parallel to the front of the instrument.
Chlorine, Total
(Method 10014
DPD Method*
*Adapted from Standard Methods for the Examination of Water and Wastewater.
Pour-Thru Cell
ULR (2 to 500 µg/L as Cl
2
)
Scope and Application: For testing trace levels of chlorine and chloramines in treated domestic and
industrial wastewater; USEPA accepted for reporting for wastewater analysis**
** U.S. Patent 5,362,650 covers the procedure. U.S. Patent 5,549,816 covers the OriFlo™ Filtration System.
1. Touch
Hach Programs.
Select program
86 Chlor. Total ULR.
Touch
Start.
2. Install the Pour-Thru
Cell and multi-
pathlength cell adapter in
the sample cell
compartment. Use the 25-
mm cell pathlength.
Note: See Section 2.6.1 on
page 26 in the Instrument
manual for installation details.
Flush with 50-mL of
deionized water.
3. Unscrew the cap from
the OriFlo™ plunger
assembly. Be sure that the
O-ring is properly seated
in the cap.
4. Install a new,
3-micron filter into the
cap well. Wet the filter
with a few drops of
deionized water.
Reassemble and hand-
tighten the cap onto the
plunger.
Tips and Techniques
DR/2400 Pour-Thru Cell Method 10014
Hach Programs

Table of Contents

Related product manuals