EasyManua.ls Logo

Hach FL1500 - Connecting to Power

Hach FL1500
374 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Install the battery near the instrument with the backup battery mounting bracket. Refer to the
documentation supplied with the mounting bracket.
Use a backup battery 3-pin half cable to connect the backup battery to the instrument. Refer to
Table 2 to connect the backup battery to the DC terminal block.
Table 2 Backup battery wiring information (AC models only)
Connection Color (cable 8307900)
12 VDC (+) White
12 VDC return (–) Black
Connect to DC power
Use a minimum 18 AWG wire to connect to DC power from a solar panel or customer-supplied DC
power.
1. Open the access door. Refer to Open the access door on page 11.
2. Put the power cable through a strain relief fitting near the DC power connector.
3. Prepare the wires.
4. Pull to remove the DC power connector.
5. Install each wire in the connector. Refer to Figure 7 and Table 3.
6. Push to install the connector in the instrument.
7. Tighten the cable strain relief.
8. Install the access door.
Figure 7 Connect to DC power
English 15

Table of Contents

Other manuals for Hach FL1500

Related product manuals