44 of 82 System Setup - 9705
KSS - March 2006 - Edition 1
Operator Manual HIAC
3.10.3 Shortening the Adapter Tube Length
The Large Volume Adapter is supplied with a 24-inch long tube. It has a two piece ferrule pre-
installed for connection into the sampling probe. The ferrule can not be re-used. If a shorter
sample tube is desired, then proceed through the following steps.
1) Leave the tube connected to the Large Volume Adapter. Cut the sampling probe end to
the desired length. Make this cut clean and square. Leave the fitting nut on the tube as
shown. Save the tubing that was cut off and observe how the ferrule is installed on it.
2) Carefully observe the loose stainless steel ferrule sleeve that has been supplied. Note
that it has a flat end and the opposite end is sharp. It is a tapered sleeve.
Fig 3-29 : Ferrule Installation
Fig 3-30 : Clean Cut of Tube