9705 - Specifications and Accessories 73 of 82
KSS - March 2006 - Edition 1
HIAC Operator Manual
Appendix B: Specifications and Accessories
B.1 Performance Specifications
o
e:
Actual flow rate is determined by the particular sensor calibration flow rate.
B.2 Accessories
Temperature Range 10 to 40°C
Relative Humidity 20 to 80%, non-condensing
Temperature Range of Sample 5 to 40°C
Viscosity Limit < 15 cp
Voltage 100 to 240 VAC, 50-60 Hz
Power 90 VA maximum
Current 0.5A at 115VAC
Fusing 1.25A, 250VAC, Type T, SB, 5 x 20mm
Dimensions (H x W x D) 40.5 x 30.1 x 34.3 cm (16.0 x 11.9 x 13.5 in)
Weight 10.7 kg (23.5 lbs)
Sample Bottle Clearance 11.8 cm (4.6 in)
Tare Volume 0.54 mL
Volume Accuracy ± 2%
Flow Rate Accuracy ± 3%
Sample Flow Rate 10 - 60 mL/min; default is 20 mL/min
Concentration Limit 10% coincidence loss @ 10,000 cts/mL (10µm PSL)
Sizing Range 1.3 µm to 150 µm
Sample Vial Sizes 5 mL, 10 mL, 20 mL, 30 mL, 50 mL, 60 mL, 100 mL, 125 mL
Software Compatibility PharmSpec 2.1 or higher
Table B-1 :
Included Accessories
Part Number Description
VP069201 Sensor Cleaning Brush
2087866-03 100 mL Adapter Ring
2087864-02 Standard Needle for 50-125 mL Bottles
2087809-01 Large Volume Adapter
210-400-517 Stylus
2087831 Needle Removal Tool