3.4 Install the sensor in the flow chamber
N O T I C E
Keep the sensor in a vertical position with the membrane down during mounting and removal. Do not shake the
sensor or oxygen can contaminate the electrolyte.
3.5 Connect the sample tubing
Note: The tube fittings and sample tubing are supplied by the user.
1. Install a ¼-inch NPT tube fitting in the sample inlet opening and sample outlet opening of the flow
chamber. Refer to Figure 3.
2. Use 6 mm (or ¼-inch) OD tubing to plumb the sample supply line to the sample inlet fitting.
3. Use 6 mm (or ¼-inch) OD tubing to plumb the sample outlet fitting to an open drain.