Installation
Note: It is good practice to use a power line filter or to connect the controller to a different branch
circuit to reduce the chance of electrical transients.
4.9.1.1 AWRS power connections
Danger
Electrocution hazard. If installed outdoors or in any wet or potentially wet location,
a Ground Fault Circuit Interrupter (GFCI) is required for the sampler. If the sampler
is installed outdoors, over-voltage protection is required.
Important Note: This product is not intended for use with a split-phase (3-wire, single-phase,
mid-point neutral) AC power system. Use this product only with a 3-wire (Hot/Neutral/Ground)
polarized AC power system. Be sure that access to the power cord plug is not blocked or obstructed.
The AWRS sampler is not provided with a power switch. The cord plug serves as the local
disconnect.
The AWRS sampler is suitable for outdoor use. The 900MAX controller gets its power
directly from the AWRS sampler.
The mains power cord for 115VAC samplers is equipped with a NEMA 5 15P plug for
standard North American outlets. The mains power cord for 230VAC samplers is
equipped with an EU 1-16P plug for standard CEE-7/7 European outlets. If you have
other types of outlets that supply compatible AC power, contact a trained electrical
technician to connect the power cord with a locally-purchased plug, in accordance
with the local safety requirements.
All AC mains power connections require a 15A (max) circuit breaker and a good earth
ground for safety and performance.
Check the power source to make sure that it satisfies the AC power requirements of
the sampler.
Figure 21 Sampler pump–time on vs. time off at 50 ºC (122 ºF)
1 Time (in minutes) when the pump must stay off 3 Time (in minutes) when the pump is on continuously
2 Example—if the pump is on for 5 minutes, the pump must stay off for 15 minutes before the next sample cycle
Goodnal STP ST041 Pre Treatment (Sigma 900 MAX All Weather Refrigerated Sampler) Vendor Manual