60
Basic programming setup
After three attempts and no sample liquid has been collected, the cycle is aborted, a
Missed Sample is reported to the Sample History Log, and the sampler begins a new
sample interval.
10-B. Press
ACCEPT to continue to Site ID.
Step 11 - Site ID
11-A. Enter a site identification number of up to 8 digits. This Site ID will appear on all data
printouts. This feature is useful when multiple sites are monitored using a single flow
meter or if data readings from multiple flow meters are collected. Press
ACCEPT.
11-B. After Site ID is accepted the following menu will appear:
11-C. If
NO, the basic setup program is complete.
Note: To return to the Advanced Sampling options at a later time, from the Main Menu select,
SETUP, MODIFY SELECTED ITEMS, PROGRAM ENTRIES.
4. Highlight Advanced Sampling and press SELECT. In the Advanced Sampling menu
highlight the an item using the
UP and DOWN keys and press SELECT.
11-D. If
YES, continue to Advanced Sampling, section 5.3.
11:00 AM 21 - APR - 01 SITE ID
ACCEPT
SITE ID:
BACKUP
00000000 CLEAR
ENTRY
(USE NUMERIC KEYPAD)
11:00 AM 21 - APR - 01 ADVANCED SAMPLING
DO YOU WISH TO
ACCESS THE ADVANCED
SAMPLING FEATURES?
YES
NO
ENTER: 1 — 999
11:00 AM 21 - APR - 01 PROGRAM ENTRIES
SELECT
INTAKE RINSES
SAMPLE RETRIES
SITE ID
ADVANCED SAMPLING
RETURN
Goodnal STP ST041 Pre Treatment (Sigma 900 MAX All Weather Refrigerated Sampler) Vendor Manual