70
Basic programming setup
18-C. Press CHANGE CHOICE to select the desired channel to trigger from. Press ACCEPT
to make the selection.
18-D. Press either
HIGH CONDITION or LOW CONDITION.
Flow Rate of Change and Rainfall are signals that only increase in value and do
not require Low Condition.
The external control signal (such as an external flow meter) must be configured in
the external equipment for the desired setpoint.
18-E. Enter the desired high or low trigger point using the numeric keypad. Press
ACCEPT
to continue.
18-F. Enter the Deadband value or, if programming for Flow Rate Of Change or Rainfall,
enter a time interval when the flow or rainfall change must take place (refer to Alarm
relays programming on page 100).
18-G. Enter the number of bottles (out of the total set) to set aside for upset samples.
These will be the last bottles in the tray.
18-H. Select the method of distribution, Sample Per Bottle or
Bottles Per Sample.
11:00 AM 21 - APR - 01 UPSET SAMPLES
UPSET CONDITION:
HIGH
CONDITION
LOW
CONDITION
11:00 AM 21 - APR - 01 UPSET SAMPLES
ACCEPT UPSET SAMPLING:
HIGH TRIGGER POINT:
CANCEL
00000 in. CLEAR
ENTRY
(USE NUMERIC KEYPAD)
11:00 AM 21 - APR - 01 UPSET SAMPLES
ACCEPT UPSET SAMPLING
DEADBAND:
CANCEL
0.000 CLEAR
ENTRY
(USE NUMERIC KEYPAD)
11:00 AM 21 - APR - 01 UPSET SAMPLES
ACCEPT UPSET SAMPLING:
NUMBER OF BOTTLES
CANCEL
2CLEAR
ENTRY
ENTER 1 — 4
Goodnal STP ST041 Pre Treatment (Sigma 900 MAX All Weather Refrigerated Sampler) Vendor Manual
Q-Pulse Id VM373
Active 29/10/2013
Page 70 of 122