70
Basic programming setup
18-C. Press CHANGE CHOICE to select the desired channel to trigger from. Press ACCEPT
to make the selection.
18-D. Press either
HIGH CONDITION or LOW CONDITION.
• Flow Rate of Change and Rainfall are signals that only increase in value and do
not require Low Condition.
• The external control signal (such as an external flow meter) must be configured in
the external equipment for the desired setpoint.
18-E. Enter the desired high or low trigger point using the numeric keypad. Press
ACCEPT
to continue.
18-F. Enter the Deadband value or, if programming for Flow Rate Of Change or Rainfall,
enter a time interval when the flow or rainfall change must take place (refer to Alarm
relays programming on page 100).
18-G. Enter the number of bottles (out of the total set) to set aside for upset samples.
These will be the last bottles in the tray.
18-H. Select the method of distribution, Sample Per Bottle or
Bottles Per Sample.
11:00 AM 21 - APR - 01 UPSET SAMPLES
UPSET CONDITION:
HIGH
CONDITION
LOW
CONDITION
11:00 AM 21 - APR - 01 UPSET SAMPLES
ACCEPT UPSET SAMPLING:
HIGH TRIGGER POINT:
CANCEL
00000 in. CLEAR
ENTRY
(USE NUMERIC KEYPAD)
11:00 AM 21 - APR - 01 UPSET SAMPLES
ACCEPT UPSET SAMPLING
DEADBAND:
CANCEL
0.000 CLEAR
ENTRY
(USE NUMERIC KEYPAD)
11:00 AM 21 - APR - 01 UPSET SAMPLES
ACCEPT UPSET SAMPLING:
NUMBER OF BOTTLES
CANCEL
2CLEAR
ENTRY
ENTER 1 — 4
Goodnal STP ST041 Pre Treatment (Sigma 900 MAX All Weather Refrigerated Sampler) Vendor Manual