calibration, but instead relates to the structure of the sludge. If you have a good / constant
sludge structure the % of homogeneity is high. The ratio between water and sludge and the
structure of the sludge is responsible for a high homogeneity value.
Q: Will the TSS portable system work correctly on low or not fully charged batteries?
A: Customers have reported issues with taking measurements, frozen menus, meters powering
off unexpectedly, and calibration issues when the batteries were low or not fully charged.
Q: Can the software be updated by the customer?
A: No – this must be sent to Service.
Q: The specifications on the meter and probe state that the acceptable low-end temperature limit is 0
degrees C. Can it be used at slightly lower temperatures?
A: We would not recommend this as it has not been tested this low and some components may
not operate correctly at this lower temperature.