Application guidelines
The battery in the refrigerator may be low after the refrigerator has been in storage for a long time.
Turn on the battery charge switch when the unit is powered up to allow the battery to be recharged.
The battery will be in full capacity after about a week’s charging.
Before loading the refrigerators, make sure the unit is at set temperature. Do not load in more than
1/3 of the storage volume to avoid thermally overloading the unit.
The display on the panel shows the sensor temperature located in the refrigerator. It is not
necessary the same as the temperature in the center of the refrigerator. The cabinet temperature will
gradually reach an equilibrium state.
Only clean the unit with light soapy water. Never use brushes, acids, gasoline, soap powders,
polishing powders or hot water to clean refrigerators as these materials may damage the interior
painting and surface, parts and components. Never wide plastic parts with volatile solvents such as
gasoline.
• If the unit is to be stored for a long period of time, turn off the power switch and the battery charging
switch.
• To reduce the possibility of temperature fl uctuation in the refrigerator, please attempt to cut down the
time to open to the door to remove and load products.
If the door is opened, the temperature of the refrigerator will warm up somewhat. That is normal.
The temperature will recover to a stable condition in a short time.
The refrigerator is designed to operate at a condition of 16 to 32 and humidity less than
85%RH . Small amount of condensation can occur on the surface of the unit if the actual condition is
outside of this range. The storage temperature of the unit, however, is not impacted. To reduce the
condensation, please improve the ventilation and drop the ambient temperature if possible.
• Only professional people should perform maintenance work.
Meaning of crossed –out wheeled dustbin:
Do not dispose of electrical appliances as unsorted municipal waste, use separate collection facilities.
Contact you local government for information regarding the collection systems available. If electrical
appliances are disposed of in landfi lls or dumps, hazardous substances can leak into the groundwater
and get into the food chain, damaging your health and well-being.
21
Specifi cation&Packing List
Specifi cation
Name Pharmaceutical Refrigerator
Model HYC-410
Exterior Dimensions
(W×D×H) (mm)
665×710×1965
Interior Dimensions
(W×D×H) (mm)
530×590×1380
Eff ective Volume 410L
Door Glass door
Insulation CFC-Free foamed-in-place urethane
Compressor Fixed Frequency Compressor
Shelves 6
Load ≤ 30kg
Refrigerating Method Force-air cooling circulation
Exterior / Interior Cold-rolled steel sheet with powder coated /HIPS plate
Condenser/Evaporater Wire and tube Condenser / Finned Evaporator
Temperature Control Microprocessor controlled
Interior light LED 1.5W
Net Weight 116kg
Voltage 220-240V~/50Hz
Rated Power/Current 350W/2.0A
Refrigerant R600a
power consumption 5.5 kW·h/24h
Alarm system
high /low temperature,High ambient temperature, Power failure,
Communication failure,Sensor fault,Low battery,Door ajar.
Battery duration for alarm
system
24 h (when the battery is fully charged)
Rechargeable Battery DC 3.6V rechargeable battery