Loading...
9
Observe the rules below for initial start-up and continuous operation:
When the refrigerator is empty, connect the power line to the appropriate dedicated
socket and make sure the power is rated voltage and frequency.
Before products go out of the factory , the battery switch is off , the users must open it
before using at the fi rst time. Once the power failure occurs, the product can realize the
function of power failure alarm.
The refrigerator has been preset to operate at 2~8 in the factory and it is not required to
set temperature.
After the power is on, the audible alarm will alarming and this is normal . Press the
“Alarm”in the control panel to stop alarming. The audible alarm continues to operate until
the temperature of sensor reaches 5 ±3 .
The temperature of refrigerator will stay at the preset temperature in a stable manner after
hours of operation.
After operation status of refrigerator is checked thoroughly, put articles into the
refrigerator.
Initial Power-on