EasyManua.ls Logo

Hallowell EMC 2002PRO - Page 20

Default Icon
22 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2/26/2018 DOCA3667B Pro Series Operating Manual.doc 20 of 22
TROUBLESHOOTING (Continued)
Symptom
Potential Causes
Possible Remedies
Ventilator sounds as though it is
No supply gas pressure.
Unkink the supply gas hose.
cycling. I can hear the valves
VOLUME control is set at it's
Replace empty tank.
clicking but nothing happens.
minimum.
Increase the VOLUME setting.
Ventilator hums with each inspiration.
Supply gas pressure at the
Unkink supply gas hose.
ventilator inlet is dropping to
Anesthesia machine power outlet
around 25 - 30 psi with O
2
flow.
is incapable of supplying the
Pipeline or tank pressure is low.
required flow. Bypass it.
Switch to new supply gas source.
Nothing happens when the
No electrical power.
Plug ventilator into the proper
ventilator is turned on. No valves
power source.
are clicking, green LED not lit.
Check voltage setting.
Check the outlet.
Check the fuse
Pop-off valve in the Bellows Base
Excessive fresh gas flow from the
Reduce the fresh gas flow.
chatters and the bellows shakes
anesthesia machine.
after it reaches the top during
exhalation.
Inspiratory flow chugs intermittently.
Pressure within the bellows
housing is exceeding 70 cm H
2
O.
Working pressure limit is
shutting off the switch Inspiratory
flow.
Inflate collapsed bellows.
Unkink the drive gas tube.
Unkink breathing system hose.
The ventilator operation sounds
normal but the TV delivered is incorrect
and or inconsistent.
Missing or damaged pop-off
valve o-ring.
Replace o-ring
Irregular cycling when using cautery or
Excessive generation of EMI
Locate and repair source of
other electro-surgical device.
and/or RFI.
disturbance.
Bellows dislodges from mounting ring.
Partially detached or improperly
Reattach or replace the bellows.
Repair defective or poorly
regulated scavenger system.
mounted bellows.
Either pressure or vacuum is
occurring at the Bellows Base
EXHAUST port.
Bellows is bulging.
P.O.V. is stuck closed.
(To determine if this is the
problem, temporarily disconnect
scavenger to isolate problem.)
Clean P.O.V. and seat.*
Everything seems normal, but the
Breathing system gas is leaking
from the system. (SEE THE
FIRST PARAGRAPH OF THIS
SECTION )
Inadequate fresh gas supply from
anesthesia machine.
Missing or damaged pop-off
valve o-ring
Hole in Bellows
Partially detached or improperly
installed bellows.
Damaged pop-off valve or valve
seat.
Anesthesia machines pop-off
valve is not completely closed.
Remove obstruction.
Increase flow.
Replace o-ring
Replace Bellows
Reattach Bellows to the mounting
ring.
Replace damaged part.
Check all tubes and tubing
connections for leaks.
Clean P.O.V. and seat.*
Close, repair or replace valve.
bellows progressively becomes less full.
*See “Cleaning the P.O.V.” page 16.