EasyManua.ls Logo

Hanna Instruments HI 902 Color - Page 207

Hanna Instruments HI 902 Color
248 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Method ID: HI1004EN
Alkalinity of Water
0-2500 mg/L CaCO
3
, pH 4.5 Endpoint
Titrator Application Methods: General Methods - r. 2.3
Description:
Method for the determination of Total (Methyl Red)
Alkalinity in water by titration of a sample to pH 4.5.
The results are expressed in mg/L (ppm) as
Calcium Carbonate.
For the determination of Phenolphthalein Alkalinity,
set the endpoint to pH 8.3.
Reference:
Standard Methods for the Examination of Water and
Wastewater 21
st
edition, Method 2320B.
Electrode:
HI 1131B Combination pH Electrode
HI 7662-T Temperature Probe
Reagents:
HI 70463 0.1N Hydrochloric Acid solution (1 L)
HI 70436 Distilled Water (1 gal)
Accessories:
HI 70300L Storage Solution (500 mL)
HI 7071 Electrode Fill Solution (4*30 mL)
HI 7004 pH 4.01 buffer solution (500 mL)
HI 7007 pH 7.01 buffer solution (500 mL)
HI 7010 pH 10.01 buffer solution (500 mL)
HI 740036P 100-mL Plastic Beakers (10 pcs)
50-mL Class-A Volumetric Pipette
Procedure:
Connect the pH electrode and temperature probe
to the titrator.
Calibrate the electrode using pH 4.01, pH 7.01
and pH 10.01 buffer. Refer to the instruction
manual for calibration procedure.
Press Select Methodfrom the main screen. Use
the arrow keys to highlight „HI1004EN Alkalinity of
Water‟ and press “Select”.
Install a 25-mL burette with 0.1N hydrochloric acid
solution (HI 70463) on pump-one and verify that
no air bubbles are present in the burette or tubing.
If necessary prime the burette until all air has
been removed.
For the determination of the exact concentration of
the hydrochloric acid follow, HI0002EN 0.1N
Hydrochloric Acid Titrant Concentration.
Use a class-A volumetric pipette to transfer
exactly 50.00 mL of sample to a clean 100-mL
plastic beaker.
Place the beaker under the stirrer assembly and
immerse the pH electrode, temperature probe and
stirrer. Ensure that the reference junction of the
pH probe is 5-6 mm below the surface.
Press “Start”. The titrator will begin the analysis.
At the end of titration, when pH 4.50 is reached,
„titration completed‟ will appear with the alkalinity
concentration. The result is expressed in mg/L of
calcium carbonate.
Remove the electrodes and stirrer from the
sample and rinse them thoroughly with distilled
water.
Record the result.
Method Parameters:
Name: Alkalinity of Water
Method Revision: 2.3
Titration Type: Standard Titration
Analog Board: Analog 1
Stirrer Configuration: Stirrer 1
Pump Configuration:
Titrant Pump: Pump 1
Dosing Type: Dynamic
min Vol: 0.050 mL
max Vol: 0.500 mL
delta E: 5.000 mV
End Point Mode: Fixed 4.500 pH
Pre-Titration Volume: 0.000 mL
Pre-Titration Stir Time: 0 sec
Measurement Mode: Signal Stability
delta E: 1.0 mV
delta t: 2.0 sec
t-min wait: 2 sec
t-max wait: 20 sec
Electrode Type: pH
Blank Option: No Blank
Calculations: Sample Calc. by Volume
Dilution Option: Disabled
Titrant Name: 0.1 N HCl
Titrant Conc.: 0.1000 N (eq/L)
Analyte Size: 50.000 mL
Analyte Entry: Fixed
Maximum Titrant Volume: 25.000 mL
Stirring Speed: 1400 rpm
Potential Range: -2000.0 to 2000.0 mV
Volume/Flow Rate: 25 mL/50 mL/min
Signal Averaging: 1 Reading
Significant Figures: XXXXX
Calculations:
Calculations: Sample Calc. by Volume
Titrant units: N (eq/L)
Titrant volume dosed: V (L)
Final result units: mg/L
Titrant conc.: 0.100 eq/L
(sample/titrant): 1.000 mol/eq
MW of sample: 100.09 g/mol
Sample volume: 50.000 mL


󰇛
󰇜


sale@ponpe.com
www.ponpe.com

Table of Contents

Related product manuals