5-4
METHODS
5.2 Standard Methods
The standard methods were developed for the most common types of analysis. Also, the
standard methods can be used as a model to create new user methods.
Only specific method parameters can be modified by the user (see Section
5.5, Method Op-
tions
section).
5.2.1 Upgrading Standard Methods
To upgrade the titrator with new standard methods, follow the steps below:
From USB Storage Device:
• Insert the USB storage device into the USB port, located on the left side of the titrator.
• Access the
General Options
screen.
• Using the
and keys, highlight the
Restore Files from USB Storage Device
option and choose .
• Using the
and keys, navigate through file types to find “standard method
files”. The list with available standard methods on the storage device will be displayed.
• Press the
or key to upgrade the titrator with the standard methods.
• Press
to return to
General Options
screen.
Note: See section 4.8 Restore Files from USB Storage Device.
From PC:
You can upgrade the titrator with standard methods from a PC using the HI 900 PC applica-
tion (see Section
4.9, USB Link with PC
).
5.2.2 Deleting Standard Methods
Unnecessary standard methods can be removed from titrator by following the procedure
below:
From
General Options
screen:
• Access the
General Options
screen.
• Using the
and keys, highlight the
Save Files to USB Storage Device
option and
press
;
• Using the
and keys, navigate through the file types to find “standard method
files”. The available standard methods will be displayed.
• Press the
or keys to remove unnecessary standard methods.
• Press
to return to the
General Options
screen.
From PC:
Unnecessary Standard Methods can be removed from the titrator using the HI 900 PC appli-
cation (see Section
4.9, USB Link with PC
).