19
MEASUREMENT PROCEDURE FOR CHLORINE ULTRA HIGH RANGE (P2)
• Turn the meter on by pressing ON/OFF. The display briefly shows all tags on.
• When the beeper sounds briefly and the LCD displays dashes, the meter is ready. The blinking
“ZERO” indicates that the instrument needs to be zeroed first.
• Fill the cuvette with 10 mL of unreacted sample, up to the mark, and replace the cap.
• Place the cuvette into the cuvette holder and ensure that the notch on the cap is positioned
securely into the groove.
• Pay attention to the selected range. To change the range, simply press RANGE/GLP .
MEASUREMENT PROCEDURE