Numeric keypad
and decimal point
Sign
Change parameter
Confirm
entry values
Call operating
parameters
Page forward in
parameter list
Datum setting
function
Status indicators
Select datum
Page backward in
parameter list
Indicator Meaning
REF Reference mark was crossed over – datum points are now stored
in non-volatile memory.
Blinking: Waiting for reference mark to be crossed over.
in. Position values displayed in inches.
1 / 2 Datum point 1 / Datum point 2 currently active.
SCL Scaling factor active.
Define workpiece edge as datum.
Blinking: Waiting for operator to confirm selection.
Define centerline between two workpiece edges as datum.
Blinking: Waiting for operator to confirm selection.
Clear entry
CL plus two-digit number:
select parameter
Clear parameter entry
Select coordinate axis
(Z-axis with ND 550 only)
Select axis-based operating parameters
Actual value and input display
(7-segment LED,
8 decades and sign)
Downwards: X-axis, Y-axis,
ND 550 only: Z-axis
MOD
0
Y
X
HEIDENHAIN
3
6
5
4
2
1
ENT
8
7
9
Z
REF
1
2
CL
1
2
SPEC
FCT
SCL
in.
Working with the digital readouts
ND 510 ND 550
HEIDENHAIN
Question and Answer IconNeed help?

Do you have a question about the HEIDENHAIN ND 510 and is the answer not in the manual?

HEIDENHAIN ND 510 Specifications

General IconGeneral
ResolutionUp to 0.1 µm
Protection classIP40
Incremental signals1 Vpp
InterfaceRS-232
Operating voltage5 V DC ±5%
Operating temperature0 to 45 °C
Storage temperature-20 °C to +70 °C
Number of axes1 to 3 axes

Summary

Datum Setting Functions

Workpiece Edge Datum

Defines a workpiece edge as the datum, considering tool diameter. Functionality confirmed by ENT press.

Centerline Datum

Defines the centerline between two workpiece edges as datum. Requires touching two edges.

Aborting Datum Setting

Procedure to cancel datum setting functions based on indicator status (blinking or steady).

Non-linear Error Compensation

Compensation Value Table Entry

Steps to select and enter compensation values for non-linear error compensation using specific keys.

Deleting Compensation Values

Procedure for deleting all entered compensation values using the CL and ENT keys.

Operating Parameters

User Parameters Overview

Describes user parameters (P00-P25) that can be changed without codes, including axis assignment.

Accessing and Modifying Parameters

Instructions for calling the parameter list, navigating, changing values, and correcting entries.

Protected Operating Parameters

Explains how to access protected parameters using code number 95148 via P00.