Technical File
File No.
Rev. No.
Non Contact Infrared Body
Thermometer
Instructions for use
Rev. Date
Page
7
vertically falling water
drops when enclosure
titled up to 15º.
Technical Specifications
Measurement Unit
°C /°F (HTD8823US)
Operating mode
Adjusted mode(Body mode)
Reference Body Site
Oral
Rated output range
34.0°C ~ 43.0°C / 93.2°F -109.4°F
Range
34.0°C ~ 43.0°C / 93.2°F -109.4°F
Accuracy
Body mode:
34.0°C ~ 34.9°C : ± 0.3°C / 93.2°F - 94.8°F : ± 0.5°F;
35.0°C ~ 42.0°C : ± 0.2°C / 95.0°F - 107.6°F : ± 0.4°F;
42.1°C ~ 43.0°C : ± 0.C 107.8°F - 109.4°F : ± 0.5°F;
Display Resolution
0.1°C /0.1°F (HTD8823US)
Three-color Backlight
(Color Alarm)
34.0°C - 35.7°C / 93.2°F - 96.3°F: no backlight;
35.8°C - 37.4°C /96.2°-99.4°F: Green (Normal Temperature);
37.5°C - 38.5°C / 99.5°F - 101.3°F: Yellow (Slight Fever);
38.6°C - 43.0°C / 101.4 - 109.4°F: Orange(High Fever)
Auto Power Off Time
≤18s
Measuring Time
Forehead mode: 2s
Forehead scan mode: 3 ~ 10s
Measuring Distance
1 CM -5CM(0.4 in -2in)
Memories
1
Power Supply Requirements
Batteries
1.5V (AAA) Alkaline batteryX2 (IEC Type LR03)
Adaptable Range
2.6V~3.6V
Environmental
Operating Condition
Operating Temperature: 5°C ~40°C (41°F ~104°F)
Relative Humidity 85% (no condensing)
Atmospheric pressure: 70 Kpa -106Kpa
Transport and Storage
Condition
Storage Temperature: - 20°C -55°C / -4 °F - 131°F,
Relative Humidity 95% (no condensing)
Atmospheric pressure: 70 Kpa -106 Kpa
Dimension and Weighting
Weight (without batteries)
47g
Size
L: 122mm X W: 30mm X H: 43mm
Compliance
Item
Compliant with