- 171 -
pressF4(OK).
5 Use fine tuning to set
the vertical angle as
V1+3, focus on the
reticle center accurately,
wait for stable value,
press [F4](OK).
[F4]
[Adjust Tilt X]
OK
HA
VA
Tilt
偏置模式 :
342°1159
F2 down 3
269°2946
91
6 After finishing, it will
display the results, press
[F4](OK), save and back
to menu.
[F4]
[Adjust Tilt X]
OK
HA
VA
Tilt
偏置模式 :
342°1159
Xk33.0859 X0-55
269°2946
100
1.4 Instrument constant setting
In Tools Menu, choose 1 Adjust, and then press [F4 to enter Const. Setting.
Press [F4](OK) to save after editing the constants.
[Const. Setting]
OK
Add Const.
Mul. Const.
偏置模式 :
-9
0
1.5 Factory setting
In Tools Menu, choose 1 Adjust, and then press [5] to enter Factory Setting.
If you need to reset the instrument parameters to factory state, you can use this
function, press key [F4] (Yes) and then the instrument will auto power off.