Page 24 199253
(6) -
EN
-
Accella™ Therapy Mattress Instructions for Use
3. Place the rolled up mattress on the top of
the bed at the head end and unroll it.
4. Check that the symbol on the
cover is at the foot of the bed.
5. Fold in two at the head end.
6. Connect the plug to the bed connector (it
clicks into place).
7. Attach the mattress using the knobs in the
center on the head section.
8. Unfold the mattress.
9. Make sure that the mattress is correctly
installed and securely attached, and in
particular that it is well centered on the sleep
surface and pushed firmly against the foot
board to avoid the danger of entrapment.
10. Connect the bed power cord to a wall outlet.
11. The mattress switches to initialization mode. A
beep sounds after 20 minutes, indicating that
the mattress is operational.
12. The status of the mattress switches to ON in
the default mode. The ALP and MCM™ modes
are active.
13. Do not place the patient on the mattress
during the initial inflation phase before the
ALP / MCM™ mode is activated. The operator must check that all the
zones of the mattress are fully inflated before installation.
14. For small patients, center the sacrum on the I-mmersion™ Therapy zone
marked out on the mattress cover.
Placing a cotton sheet on the mattress will improve patient comfort and make
it easier to administer care.
It is advisable not to place any waterproof materials (e.g., drapes) on
mattresses with the MCM™ function, as doing so would reduce the
performance of the function.