295
────────────────────────────────────────────────────
13.4 Storing, Retrieving and Deleting Data on the Floppy Disk or PC Card
────────────────────────────────────────────────────
NOTE
Example
NOTE
・Data stored in the text or BMP format is not readable by the 8835-01.
・Only the waveform data and setup files created by the 8835-01 can be read.
・(*) When batch saving with the Recorder & Memory, the measurement data
(with REC, MEM extensions) are created together with an index file. When
only the measurement data are read, these are read to the respective
functions. To read to the Recorder & Memory, read the index file.
・(*) Do not change or delete the name of a file to be batch read. The batch
reading may not be executed correctly. (*): Advanced version
・In case of "overwrite load, the settings specified in the 8835-01 take
precedence over other settings (see "File contents" in Section 13.4.5).
Auto setup function
When the power is switched on with a floppy disk or PC card inserted, the
setting up of the 8835-01 is performed automatically by reading the setting
data file called STARTUP. SET on the floppy disk.
The same measurement conditions can be simply established by using this
function.
Method
Define the setup data file name as STARTUP.SET, and save the file to the
root directory of the floppy disk or PC card. The file is read to the 8835-01
when the equipment is turned on.
When STARTUP.SET is in both floppy disk and PC card, file is read from
floppy disk.