Chapter 3 Specifications
32
Model IR4030
Rated measurement
voltage (DC)
25 V 50 V 125 V
Maximum effective
indicated value
20 M
Center scale value
0.5 M
1st effective
measuring range
0.02 M to 10 M
2nd effective
measuring range
0.005 M to 0.02 M
10 M to 20 M
Lower limit measurement
resistance value to be
maintained rated
measurement voltage
0.025 M 0.05 M 0.125 M
Continuous operating time
Approx. 20 hours (no load)
Input error protection
for 10 seconds
150 VAC
Model IR4031
Rated measurement
voltage (DC)
50 V 125 V 250 V
Maximum effective
indicated value
100 M
Center scale value
2 M
1st effective
measuring range
0.1 M to 50 M
2nd effective
measuring range
0.01 M to 0.1 M
50 M to 100 M
Lower limit measurement
resistance value to be
maintained rated
measurement voltage
0.05 M 0.125 M 0.25 M
Continuous operating time
Approx. 20 hours (no load)
Input error protection
for 10 seconds
600 VAC