Troubleshooting
When a warning occurs, a message will be displayed for several seconds.
No. Message Solutions
NTP client disabled Set the NTP client function in the NTP
settings to [ON].
See “Synchronizing the time” in the
Instruction Manual.
No server address entered Enter the server address in the NTP
settings.
See “Synchronizing the time” in the
Instruction Manual.
Invalid key The pressed key is disabled, for example
because measurement is in progress. Press
the key after the operation completes.
Modication not available while measuring Press the STOP key to stop measurement
and then change the setting.
Settable range exceeded The entered value does not fall within the
valid setting range.
Enter an appropriate value.
Pre-trigger time and settings modied Changing the recording interval or recording
time may shorten the time that can be set
for the pre-trigger.
Check the updated pre-trigger settings.
Trigger or alarm setting value of this channel
modied
The trigger or alarm settings have changed.
Check the updated settings.
See “2.2 Enabling the Trigger Function” and
“4.1 Conguring Alarms” in the Instruction
Manual.
Burn-out settings modied The wire break detection function has been
set to [OFF].
To use the wire break detection function,
either increase the recording interval or
reduce the number of channels being
measured.
See “1.3 Setting Measurement Conditions”
in the Instruction Manual.
Cannot activate Burn Out (broken-wire)
detection. Set the unit update interval to
automatic or a rate slower than current
settings.
The user-specied setting does not support
the wire break detection.
Set the data refresh interval to [Auto] or an
interval longer than that presently set.
See “1.3 Setting Measurement Conditions”
in the Instruction Manual.
File partition period modied Changing the recording interval will cause
the segment time to change. Check if the
settings are appropriate.
Recording interval modied The measurement preferences of all
channels of the High Speed Voltage Unit
and Strain Unit have been set to [OFF],
forcing the recording interval setting to
change. Check the updated recording
interval.
Settings for external I/O3 modied to
external trigger
The [External input 3] terminal was set to
[Trigger input] because the external trigger
was set to [ON]. Check if the settings are
appropriate.
See “2.5 Applying Triggers Based on
External Sources” in the Instruction Manual.