130-0047 Issue 8
G1010OEM OXYGEN ANALYSER
HANDBOOK.
1.0 GENERAL DESCRIPTION
The G1010 is a microprocessor controlled oxygen analyser which
provides a performance and a range of features without parallel for
an analyser of this type and cost.
IMPORTANT NEW SENSOR IDENTIFICATION
In order to make sensor ordering easier we have renamed
our galvanic sensors used in these instruments into 3
groups:
N type - (formally C/S & C/N)
L type - (formally LL or C/NLL)
This handbook reflects these changes; if you have any
queries concerning sensor configuration, then please
contact us with the instrument serial number, and we will be
pleased to assist.
Page 1 of 18
2.0 SPECIFICATION
Display
Multi digit LCD - character height 12.7mm
Display ranges
1) % version (N type sensor) - Display range 0 to 55%. Resolution:
0.1% from 10% to 25%; 0.01% below 10%
2) ppm version (L type sensor) - Display range 0ppm to 5%
Stability
Better than 2% of full-scale per month
Sensor life
Up to 18 months
Sample flow
Between 100 and 300 ml/min for optimum operation. Max. 1 l/min
Sample pressure
The pressure applied to the sensor is determined by the vent
pressure which should be atmospheric for quoted accuracy.
Note: The sensor must not be subjected to rapid pressure changes.
Sample temperature
-5 to +40 °C (non condensing)
Sample connections
Inlet and outlet: captive seal compression fittings suitable for 0.25
inch (or 6mm) o/d tube
Page 2 of 18