HOBO
®
U12 Temp/RH/Light/
External Data Logger
(Part # U12-012)
© 200 Onset Computer Corporation
Part #: MAN-U12-012, Doc #: 7661-C
Thank you for purchasing a HOBO data logger. With proper care, it will give you years of accurate and reliable measurements.
The HOBO U12 Temperature/Relative Humidity/Light/External Data Logger is a four-channel logger with 12-bit resolution and can record up to
43,000 measurements or events. The external channel accepts a variety of sensors, including temperature, and split-core AC current sensors as well
as 4-20 mA and voltage input cables (sold separately). The logger uses a direct USB interface for launching and data readout by a computer.
An Onset software starter kit is required for logger operation. Visit www.onsetcomp.com for compatible software.
Inside this package:
HOBO U12 Temp/RH/
Light/External Data Logger
Mounting kit with magnet, hook and loop tape, tie-wrap mount, tie wrap, and two screws.
Specifications
Measurement range
Temperature: -20° to 70°C (-4° to 158°F)
RH: 5% to 95% RH
Light intensity: 1 to 3000 footcandles (lumens/ft
2
) typical; maximum value varies from 1500 to 4500 footcandles (lumens/ft
2
)
External input channel (see sensor manual): 0 to 2.5 DC Volts
Accuracy
Temperature: ± 0.35°C from 0° to 50°C (± 0.63°F from 32° to 122°F), see Plot A
RH: +/- 2.5% from 10% to 90% RH (typical), to a maximum of +/- 3.5%. See Plot B.
Light intensity: Designed for indoor measurement of relative light levels, see Plot D for light wavelength response
External input channel (see sensor manual): ± 2 mV ± 2.5% of absolute reading
Resolution
Temperature: 0.03°C at 25°C (0.05°F at 77°F), see Plot A
RH: 0.03% RH
Drift
Temperature: 0.1°C/year (0.2°F/year)
RH: <1% per year typical; RH hysteresis 1%
Response time in airflow of
1 m/s (2.2 mph)
Temperature: 6 minutes, typical to 90%
RH: 1 minute, typical to 90%
Time accuracy ± 1 minute per month at 25°C (77°F), see Plot C
Operating temperature
Logging: -20° to 70°C (-4° to 158°F)
Launch/readout: 0° to 50°C (32° to 122°F), per USB specification
Battery life 1 year typical use
Memory 64K bytes (43,000 12-bit measurements)
Weight 46 g (1.6 oz)
Dimensions 58 x 74 x 22 mm (2.3 x 2.9 x 0.9 inches)
The CE Marking identifies this product as co
mplying with all relevant directives in the European Union (EU).
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Question and Answer IconNeed help?

Do you have a question about the Hobo U12-012 and is the answer not in the manual?

Hobo U12-012 Specifications

General IconGeneral
BrandHobo
ModelU12-012
CategoryData Loggers
LanguageEnglish

Related product manuals