100 Annex
Guidance and manufacturer’s declaration – electromagnetic immunity
The point 2 ventilator is intended for use in the electromagnetic environment specified below.
The user of the point 2 ventilator should ensure that it is used in such an environment.
Immunity tests IEC 60601
Test level
Compliance level Electromagnetic
environment –
guidance
Electrostatic
discharge acc. to
(ESD) IEC 61000-4-2
± 8 kV
contact
± 2, 4, 8, 15 kV
air
± 8 kV
contact
± 2, 4, 8, 15 kV
air
± 2, 4, 8, 15 kV
air display
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Radiated RF
disturbances acc. to
IEC 61000-4-3
10 V/m
80 MHz – 2.7 GHz
80% AM at 1kHz
10 V/m
Portable and mobile
communications
equipment (including
cables and antennas)
should be used no
closer to any part of the
Point 2 home ventilator
than the recommended
separation distance of
0.3 m.
27 V/m
385 MHz PM: 18Hz
27 V/m
28 V/m
450 MHz
FM ± 5Hz: 1kHz sinus
28 V/m
9 V/m
710, 745, 780 MHz
PM:217Hz
9 V/m
28 V/m
810, 870, 930 MHz PM:
18Hz
28 V/m
28 V/m
1720, 1845, 1970 MHz
PM: 217Hz
28 V/m
28 V/m
2450 MHz PM:217Hz
28 V/m
9 V/m
5240, 5500, 5785
PM:217Hz
9 V//m